Manufacturer:
Endress+Hauser
Product Name:
FMU862
Product Description:
Ultrasonic
Profile:
HART EDD Only
More Information:
Ultrasonic Level Transmitter (Not available anymore, Successor Prosonic S)
Prosonic FMU 860...862 Level and flow measurement with ultrasonics.
Simple to start up, easy to use, flexible field mounted ultrasonic
transmitters
Areas of application
Prosonic FMU 860...862 ultrasonic transmitters are for use with the
Prosonic FDU 80....86 family of sensors.
The transmitter determines levels in silos and tanks and calculates the
volume of solids and liquids they contain.
With the appropriate certified sensor, measurements can be made in
explosion hazardous areas, zone 1 or in combustible dusts.
When used for applications in fresh and waste-water, Prosonic measures
* flow rates in flumes and weirs
* water levels
and can control screen cleaning and pumps.
Benefits at a glance
A customised instrument programme for the specific application choice of
transmitters
* For the field or control room
* Single or two-channel versions with three or five relays, also with
optional totaliser
* With optional serial interface for remote configuration (HART
protocol)
* With PROFIBUS-DP or RS-485 interface
Intelligent commissioning, ultrasonic measurement par excellence
* Instrument parameters arranged in a simple matrix
* Various linearisation functions, totalisers, and all common Q/h
curves on call
* Quick commissioning and stable measurement thanks to signal pattern
recognition by fuzzy logic and the application parameter.
The most important technical data
Input:
Sensors, separate switch input, separate temperatur sensor
Output:
0/4...20 mA galvanically separated, three or five inpendent relays, each
with one potential-free change-over contact
Profibus-DP
Power supply:
AC 230 V or 115 V or 48 V or 24 V
DC 20...30 V
Certificates:
Standard, CSA
FieldComm Group (FCG) makes available to its members and to the public certain device driver (DD) files, in various forms, that have passed certain tests conducted by or through FCG. The DD files are, or are created from, submissions to FCG by FCG members and are provided by FCG AS IS. FCG disclaims all liability for the accuracy, suitability or use of the DD files. Each member creating DD files has its own policies regarding supporting use of those files, and may, for instance support use of DD object code files, but not DeviceInfo or other files. While FCG is under no obligation to provide support for use of any DD files, FCG welcomes all inquiries on the subject and will often be able to assist.