Manufacturer:
Emerson Automation Solutions
Product Name:
Rosemount Analytical 1066-P-FF
Product Description:
pH, ORP, Redox Transmitter
Profile:
FOUNDATION Field Device
More Information:
The 1066-P-FF represents the latest generation in a long line of Emerson
Rosemount Analytical Fieldbus pH transmitters, going back to the world's
first, introduced in 1998.
The 1066-P-FF measures pH (0 to 14 pH) and ORP/Redox (-1,500 to 1,500
mV), and provides full comprehensive sensor, transmitter, and calibration
diagnostics, which are available to the bus via Field Diagnostics. The
Common Software Download Specification allows software upgrades to be
downloaded to the transmitter, while it is in service. As has always
been the case with our transmitters, the 1066-P-FF is provided with a
Link Active Scheduler.
4 AI Function Blocks are provided for PV, temperature, reference
electrode impedance, and glass electrode impedance. An AO Function Block
is provided to allow the transmitter to use a temperature measurement
from the bus for temperature compensation.
Additional Function Blocks are provided for use by the 1066-P-FF or other
transmitters on the segment, and include:
Math Function Blocks: Arithmetic and Integrator
Analog Control Function Blocks: PID Control, Input Selector, Signal
Characterizer, Control Selector, and Output Splitter.
The 1066-P-FF transmitter in housed in a NEMA 4X, ½ DIN enclosure. The
local user interface consists of a large, monochromatic LCD display,
which provides an intuitive menu, accessed using 4 function keys and 4
selection keys. The 1066-P-FF is available in a FISCO compliant version.
Manufacturer ID (hex):
524149
Device Type ID (hex):
4089
DEV_REV (hex):
01
DD Revision:
0x05, 0x05(DD5), 0x03(CF)
FOUNDATION Physical Layer Profile:
111 - Standard-power signaling, bus-powered, Entity Model I.S.
113 - Standard-power signaling, bus-powered, non I.S.
511 - Low-power signaling, bus-powered, FISCO I.S.
Stack Manufacturer:
Emerson Process Management, LLLP
FOUNDATION Stack Class:
31PS, 32L
Stack Version:
TH6.0
Registered Blocks:
1-RB(e), 4-AI(e), 1-AO(e), 1-PID(e), 1-IS(e), 1-AR(s), 1-IT(s), 1-SC(s)
Other Blocks:
1-CS(s), 1-OS(s),<br>1-MainTB(c), 1-FactTB(c)
Registered Features:
Alarms and Events
Function Block Linking
Trending
Multi-bit Alert Reporting
Block Instantiation
Common Software Download
Field Diagnostics
Registration Authority:
FieldComm Group
Registration Status:
Registered
Tester Version:
6.0.1
DD/CFF Tester Version:
6.1.0